Description of operational conditions for ICP-MS of fractionated materials.

OPERATIONAL CONDITIONS


RF power (W)

1550


Plasma gas flow rate (L/min)

15


Hydrogen flow rate (mL/min)

2.5


Helium flow rate (mL/min)

2.5


Carrier flow rate (L/min)

0.9


Make-up gas (L/min)

0.1


Auxiliary gas (L/min)

0.9


Sample flow rate (mL/min)

0.4


Acquisition time per isotope (sec)

0.1


Repetition

3


Total acquisition time for 19 isotopes (sec)

2.85


Total running time (sec)

1880


Tuning solution:

% RSD

< 3%


Sample nebulizer tubing:

Material

Tygon

Internal diameter (mm)

1.02


AF4 carrier tubing:

Material

PEEK

Internal diameter (mm)

0.25